Product description and applicable:
HPCS6500 high-precision photometirc,Colorimetric and Electric Test System, the whole machine adopts aluminum alloy chassis, embedded high-speed processor, 10.1-inch color LCD Touch screen, Use interaction. The system integrating high-precision spectrometer, DC power supply,AC power supply, Digital power meter in one device, use independently also can connect to pc software.
Technical features:
1. The system design and measurement results fully comply with the relevant standards of CIE (International Commission on Illumination)
2. Can realize LED current characteristics test, luminous flux test and external trigger test (optional)
3. Precision DC stabilized voltage and constant current power supply
4. DC power supply with high precision and high resolution can reach 1mV/0.1mA; constant voltage/constant current automatic switching
5. Equipped with a 10.1-inch screen and high-sensitivity full-screen capacitive touch, it is to enhance the interactive experience and facilitate user use.
6. Equipped with PC software, online measurement, real-time analysis, file saving, and test reports can be exported at one time to facilitate users to share and communicate test results.

High-precision spectrum analyzer
1. Wavelength range: 380nm-780nm
2. Wavelength accuracy: ±0.3nm
3. Optical platform: telephoto cross-asymmetric CT spectroscopy system
4. Optical bandwidth: 2nm (FWHM)
5. Detector: HAMAMATSU (Hamamatsu, Japan) back-illuminated precision array CCD
6. Integration time: 10uS-10S
7. Chromaticity coordinate accuracy: ±0.001 (under standard A light source)
8. Chromaticity coordinate repeatability: ±0.0003 (under standard A light source)
9. Luminous flux measurement range: 0.1lm-200000lm (with different integrating spheres)
10. Luminous flux accuracy: Level 1
11. Photometric linearity: ±0.5%
Precision DC regulated constant current power supply
1. Output range (voltage / current): 0-60V / 0-5A
2. Measurement accuracy (voltage/current): ±(0.02% reading +5mV) (25±5°C) / ±(0.05% reading +5mA) (25±5°C)
3. Resolution (voltage/current): 0.01V/0.1mA
4. Ripple and noise (voltage/current): ≤1mVrms / ≤6mArms
Precision AC variable frequency test power supply
1. Power input range: 220V±5%, 50/60Hz
2. Output power: 750W
3. Output voltage: low voltage range 1.0V-150V, high voltage range 151.0-300.0V
4. Output current: high voltage 0-2.5A, low voltage 0-5.0A
5. Voltage distortion: ≤2% (THD)
Harmonic electrical parameter measuring instrument
Measuring range: voltage 3.0-300V; current 0.005-20.0A
Measurement accuracy: ± (0.4% reading + 0.01% range + 1 word)
Harmonic function: Voltage and current total harmonics and 2-50th harmonic values
Measurement parameters:

Rich software features: